See The Future
Princeton Instruments
Scientific Imaging Industrial Imaging Spectroscopy X-Ray Acton Optics
Search

Transmission Spectroscopy

Incident light interacting with a sample experiences reflection and absorption losses (Fig.1) which are minimized when the incidence angle qi = 0. Standard measurements of transmission are performed using normal incidence when the incident light beam is positioned strictly perpendicular to a sample surface.

         

                    Figure 1. Light transmission in the presence of 
                    Reflection and Absorption losses.

Transmittance

Transmittance, or the coefficient of transmission, is a portion of incident light at a certain wavelength l passing through a sample:

Tl = I/I0,

Where I0 is the intensity of the incident light at l; I is the intensity of transmitted light at l.

For practical convenience, transmittance is usually expressed in percentage:

Tl% = I/I0  100%

Transmittance has close relations with Absorbance (in the absence of Reflection):

Al = 2 – log10 (Tl%) = - log10 (Tl)

According to the energy conservation law, the total energy of reflected, absorbed, and transmitted beams equals the energy of the incident beam.


Recommended Products:

Acton Series Monochromators & Spectrographs
-  Positrak™ grating stabilization offers simple calibration
-  Optimized coatings for higher throughput
-  Interchangeable grating turrets with a wide selection of gratings


PIXIS CCD Cameras
-  Lifetime vacuum guarantee for worry-free operation
-  Deep cooling without the need for liquid circulators
-  Up to 1000 spectra/sec data acquisition

SPEC-10 CCD Cameras
-  Exclusive dual amplifier design offers high SNR or high dynamic range
-  Superior low light sensitivity through deeper cooling

PI-MAX ICCD Cameras
-  5MHz digitization provides up to 15 frames per second
-  500ps gating for ultra fast time resolution
-  Exclusive UNIGEN photocathode for highest sensitivity from deep UV to NIR

OMA V InGaAs Array Cameras
-  Excellent NIR sensitivity for demanding spectroscopy applications
-  Electronic shutter allows intergration times down to 20 µsec
-  1MHz digitization provides up to 1800 spectra/sec